xray quality assurance raydent meditech

X-Ray Quality Assurance Services

Quality Assurance is very important for radiation safety of Doctors, Patients and all the hospital staff working with diagnostic machines & it is mandatory for all new and old machines according to rules set by AERB ie Atomic Energy Regulatory board.Quality Assurance is a check of your radiological diagnostic machines parameters against permissible & safe limits as set by AERB.

Anyone using Radiological Diagnostic machines must have a Elora Registration, Quality Assurance certificate from a AERB licensed vendor  and a proper license issued by AERB, Only then one is allowed to operate Radiological Diagnostic Machines.

Raydent Meditech Provide AERB Approved Radiological Quality Assurance Services of Diagnostic X-Ray Machines all over India.

How To Get Your X-Ray Machine Quality Assured ?

We Provide a Hassle Free, Swift & Low Cost Quality Assurance Services for X-Ray Machines, Here are few steps Involved:

  • Quotation — We Provide you a competitive price for QA Service
  • Documents — You Must Provide the required documents regarding your machine & centre.
  • Registration —  Registration on ELORA online application provided by AERB
  • Engineer Visit — Our AERB Certified Service engineer visits your centre and perform QA Tests
  • Report —  We Send you a Pass/Fail report according to the data collected by our engineer as per AERB Guidelines.
  • Upload & License —  You can upload your QA Report over online ELORA Application and get a license issued from AERB online.

We Conduct Following Tests to check the Quality of Diagnostic X-Ray Machines as set by AERB:

  • Accuracy of Operating Potential (kVp).
  • Accuracy of Timer.
  • mA/mAs Linearity (CoL).
  • Reproducibility of output (CoV).
  • Low Contrast And High Contrast Resolution.
  • Total Filtration.
  • Radiation Leakage level from X-Ray tube housing.
How can we help you?

Contact us at the Raydent Meditech office or submit a inquiry online.